Industriële thermometer TM131 met massieve thermowell
Overzicht specificaties
Accuracy
Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
depending on configuration
Max. process pressure (static)
depending on the configuration up to 500 bar
Operating temperature range
PT100 TF StrongSens: -50 °C ...500 °C (-58 °F ...932 °F) PT100 QuickSensTF: -50 °C …200 °C (-58 °F …392 °F) PT100 WW: -200 °C ...600 °C (-328 °F ...1.112 °F) PT100 TF: -50 °C ...400 °C (-58 °F ...752 °F) Typ K: max. 1.100 °C (max. 2.012 °F) Typ J: max. 800 °C (max. 1.472 °F) Typ N: max. 1.100 °C (max. 2.012 °F)
Modulaire industriële RTD- of TC-thermometer, compleet met massieve thermowell of voor gebruik met thermowell ter plaatse. Leverbaar met Endress+Hauser-temperatuurtransmitters voor betere nauwkeurigheid en betrouwbaarheid - eenvoudige customizing door het kiezen van de uitgangen en communicatieprotocollen. Optioneel leverbaar met bluetooth®-connectiviteit. Thermometer met de nieuwste innovaties zoals iTHERM QuickNeck en temperatuursensortechnologieën als iTHERM QuickSens en iTHERM StrongSens.
Universeel toepassingsgebied
Meetbereik: –200 tot +1100 °C (–328 tot +2012 °F)
Drukbereik tot 500 bar (7.252 psi)
Trillingsbestendige sensorelementen tot 60 g
Verbeterde onderhoudsvriendelijkheid (vervangen van sensor zonder procesonderbreking), eenvoudige en veilige herkalibratie van het meetpunt
Voordelen
Tweede procesafdichting met storingsindicatie geeft waardevolle informatie over de status van het instrument
iTHERM QuickSens: zeer korte responstijden van 1,5 s voor een optimale procesbewaking
iTHERM StrongSens: onovertroffen trillingsbestendigheid (> 60 g) voor maximale fabrieksveiligheid
iTHERM QuickNeck – bespaart tijd en geld dankzij eenvoudige herkalibratie zonder gereedschap
Bluetooth®-connectiviteit (optioneel)
Internationale certificaten: explosieveiligheid volgens ATEX, IECEx, CSA C/US en CCC
Iplom SpA sought to optimize its refining process to extend catalyst life and to increase safety in a hydrodesulfurization (HDS) application through precise 3D temperature profiling.
metric style modular temperature assembly universal range of application suitable for hazardous areas suitable for high process pressures with neck or QuickNeck incl. thermowell / protection tube (metal) or to be installed into thermowell can be used with StrongSens
Thermowell / protection tube
barstock (drilled) or to be used with thermowell
Insert / probe
mineral insulated (MI), flexible
Outer diameter protection tube / Insert
16 mm (0,63") 18 mm (0,71") 19 mm (0,75") 19,5 mm (0,77") 20 mm (0,79") 22,2 mm (0,87") 24 mm (0,95") 25 mm (0,98") 25,4 mm (1,0") 26 mm (1,02") 27 mm (1,06") 30 mm (1,18") 33,4 mm (1,31")
PT100 TF StrongSens: -50 °C ...500 °C (-58 °F ...932 °F) PT100 QuickSensTF: -50 °C …200 °C (-58 °F …392 °F) PT100 WW: -200 °C ...600 °C (-328 °F ...1.112 °F) PT100 TF: -50 °C ...400 °C (-58 °F ...752 °F) Typ K: max. 1.100 °C (max. 2.012 °F) Typ J: max. 800 °C (max. 1.472 °F) Typ N: max. 1.100 °C (max. 2.012 °F)
Max. process pressure (static)
depending on the configuration up to 500 bar
Accuracy
Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
depending on configuration
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvals
ATEX II3G Ex ec IIC T6, II3D CSA C/US General Purpose CSA C/US IS, NI I/1+2/A-D CSA C/US XP, NI, DIP I, II, III/1+2/A-G CSA C/US DIP II, III / 1/E-G EAC 0Ex ia IIC T6...T1 Ga X EAC Ga/Gb Ex d IIC T6...T1 X EAC Ex ta/tb IIIC 85oC...450oC Da/Db X EAC Ga/Gb Ex ia IIC T6...T1 X EAC 0Ex ia IIC T6...T1 Ga X, Ex ia IIIC 85oC...450oC Da/Db X EAC Ga/Gb Ex d IIC T6...T1 X, Ex ta/tb IIIC 85oC...450oC Da/Db X INMETRO Ex ia IIC T6 Ga/Gb, Ex ia IIIC Da/Db INMETRO Ex d IIC T6 Ga/Gb, Ex tb IIIC Da/Db INMETRO Ex nA IIC T6 Gc CCC Ex ia IIC T6 Ga/Gb CCC Ex d IIC T6 Ga/Gb, Ex tD A20/A21 CCC Ex tD A20/A21 CCC Ex nA IIC T6 Gc UK II1/2G Ex ia IIC T6 Ga/Gb, II1/2D Ex ia IIIC Da/Db UK II3G Ex nA IIC T6 Gc, II3D Ex tc IIIC Dc UK II1/2G Ex db IIC T6 Ga/Gb, II1/2D Ex ta/tb IIIC Da/Db ATEX IECEx II1/2D Ex ta/tb IIIC Da/Db ATEX IECEx II1/2G Ex ia IIC T6 Ga/Gb ATEX IECEx II1/2G Ex ia IIC T6, II1/2D ATEX IECEx II1/2G Ex db IIC T6, II1/2D
Temperatur und Systemprodukte für
Kraftwerke & Energie, Chemie und Öl & Gas
PU (publicatie)
Duitse versie - 06/2023
New version available in English
Temperatur und Systemprodukte für Kraftwerke & Energie, Chemie und Öl & Gas Enthält iTHERM ModuLine, iTHERM MultiSens, iTEMP Temperaturtransmitters, Systemprodukte
Temperature and system products for
Power & Energy, Chemical and Oil & Gas
including iTHERM ModuLine,
iTHERM MultiSens,
iTEMP temperature
Transmitters, system products
PU (publicatie)
Engelse versie - 06/2023
New version available in English
Temperature and system products for Power & Energy, Chemical and Oil & Gas including innovative iTHERM ModuLine, iTHERM MultiSens temperature assemblies, iTEMP temperature transmitters and system products
WirelessHART communication interface Transmission rate: Nominal 250 kBits/s Operating frequency: 2.4 GHz (ISM band) Transmission range under reference conditions: Outdoor 250 m, indoor 50 m
WirelessHART communication interface Transmission rate: Nominal 250 kBits/s Operating frequency: 2.4 GHz (ISM band) Transmission range under reference conditions: Outdoor 250 m, indoor 50 m
We maken gebruik van cookies om uw browse-ervaring te verbeteren, statistische gegevens te verzamelen om de site-functionaliteit te optimaliseren en om reclame of content op maat te presenteren.
Door "Accepteer alles" te selecteren gaat u akkoord met ons gebruik van cookies.
Bekijk voor meer details onze cookie-richtlijnen .