Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
t90 starting at < 1,5 s iTHERM QuickSens depending on configuration
Max. process pressure (static)
depending on the configuration
Operating temperature range
PT100 TF iTHERM StrongSens: -50 °C ...500 °C (-58 °F ...932 °F) PT100 TF iTHERM QuickSens: -50 °C …200 °C (-58 °F …392 °F) PT100 WW: -200 °C ...600 °C (-328 °F ...1.112 °F) PT100 basic TF: -50 °C ...200 °C (-58 °F ...392 °F) Typ K: max. 1.100 °C (max. 2.012 °F) Typ J: max. 800 °C (max. 1.472 °F) Typ N: max. 1.100 °C (max. 2.012 °F)
iTHERM ModuLine TM112 is an easy-to-use imperial direct contact modular thermometer, featuring outstanding RTD or TC sensor technology. It is suitable for every application, from basic functionality to high-end performance in the chemical, oil & gas and power & energy industries.
For universal use
Measuring range: –200 to +1 100 °C (–328 to +2 012 °F)
Pressure range: up to 75 bar (1 088 psi)
Voordelen
User-friendly and reliable from product selection to maintenance
iTHERM inserts: full traceability and consistently high product quality for reliable measured values
iTHERM QuickSens: fastest response times of 1.5 s for optimum process control
Aangezien de temperatuur van gecomprimeerd waterstof tijdens het vullen toeneemt door expansie, is temperatuurmeting om veiligheidsredenen essentieel bij waterstoftankstations.
Onze niet-invasieve thermometer kan eenvoudig achteraf op een bestaande leiding worden gemonteerd en in het regelsysteem worden geïntegreerd om een efficiënte stand-bywerking van een stoomturbine mogelijk te maken.
De niet-invasieve thermometer iTHERM SurfaceLine TM611 meet de temperatuur nauwkeurig en snel, waardoor de prestaties en efficiëntie van de warmtewisselaar worden geoptimaliseerd.
The Ziegler-Natta process uses hazardous chemicals like titanium tetrachloride. Precise temperature measurement is crucial to prevent runaway reactions and ensure safety and efficiency.
imperial style universal range of application suitable for hazardous areas can be used with iTHERM StrongSens, iTHERM QuickSens direct process contact
Thermowell / protection tube
without, direct process contact
Insert / probe
mineral insulated (MI), flexible
Outer diameter protection tube / Insert
Insert: 1/8" 1/4" 3/8" 3/8" reduziert auf 3/16" 3,0 mm 6,0 mm
Max. immersion length on request
up to 180"
Material protection tube/ thermowell
Insert Material: 316L (1.4404) Alloy 600 (2.4816)
Process connection
Thread: NPT1/2", NPT3/4" Compression fitting, also spring load: NPT1/4", NPT1/2"
Tip shape
straight stepped
Surface roughness Ra
0,76 μm (30 μin.) 1,6 μm (63,0 μin.)
Operating temperature range
PT100 TF iTHERM StrongSens: -50 °C ...500 °C (-58 °F ...932 °F) PT100 TF iTHERM QuickSens: -50 °C …200 °C (-58 °F …392 °F) PT100 WW: -200 °C ...600 °C (-328 °F ...1.112 °F) PT100 basic TF: -50 °C ...200 °C (-58 °F ...392 °F) Typ K: max. 1.100 °C (max. 2.012 °F) Typ J: max. 800 °C (max. 1.472 °F) Typ N: max. 1.100 °C (max. 2.012 °F)
Max. process pressure (static)
depending on the configuration
Accuracy
Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
t90 starting at < 1,5 s iTHERM QuickSens depending on configuration
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvals
ATEX Ex ec ATEX IECEx Ex ia CSA C/US IS, NI, XP, DIP INMETRO Ex ia