Zoek op trefwoord, order- of productcode of serienummer, bv. “CM442” of “Technische informatie”
Voer tenminste 2 karakters in om het zoeken te starten.
  • Geschiedenis
product picture of the process gas analyzer MCS300P

MCS300P
process gas analyzer

Simultaneous process monitoring of up to 6 measuring components

IMCS3P
Please Wait (spinning wheel)
Prijzen worden geladen
Prijzen tijdelijk niet beschikbaar
Price on request
Bezig met activeren van prijs
Prijs: Informatie verstrekken
from CHF ??.-
Filter meting-/ toepassingstaak
Meet-/toepassingstaak

Configureer om de CAD-tekeningen van dit product te downloaden

  • Meetprincipe

    Interference filter correlation, Gas filter correlation

  • Measured variables

    Br2, CCl4, C2Cl4, CCl2F2, C3F6, CH4, C2H2, C2H4, C2H6, C3H4, C3H6, C3H8, C4H6, C4H8, C4H10, C6H6, C6H14, C7H8, CH3Cl, CH2Cl2, CHCl3, C2H3Cl, C2H2Cl2, C2H4Cl2, C2HCl3, C6H5Cl, C6H4Cl2, CHCl2F, CH3OH, C3H7OH, C2H4(OH)2, CH3CHO, HCOOH, CH3COOC2H3, C2H2O, Cl2, CO, CO2, COCl2, HCl, HCN, HF, H2O, NH3, NO, N2O, NO2, R-NCO, SO2, SiF4, many other IR / VIS active gases and liquids

  • Measuring range

    More than 60 measuring components available (depending on concentration and sample gas composition)
    Up to 6 components simultanously
    2 measuring ranges per component
    Automatic measuring range switching (adjaustable)
    2 limit values per component
    Measuring ranges depend on application and combination of measuring components

  • Ambient temperature range

    +5 °C ... +40 °C
    Temperature change maximum ±10 °C/h

  • Ambient humidity

    ≤ 80 %
    Non-condensing

  • Hazardous area approvals

    IECEx:
    Ex pxb IIC T3...T4 Gb
    Ex pzc IIC T3...T4 Gc
    Process cuvette: Ex eb IIC Gb
     
    ATEX:
    II 2G Ex pxb IIC T4 Gb
    II 2G Ex pxb IIC T3 Gb
    Process cuvette: II 2G Ex e IIC Gb
    II 3G Ex pzc IIC T4 Gc
    II 3G Ex pzc IIC T3 Gc

  • Degree of protection

    IP65: measuring cuvette depending on version

  • Product safety

    CE